Wednesday, November 7, 2018
20181102 Site calibration and Site brought back to campus
Performed Calibration before bringing back instrument back to campus due to end of measuring season.
Friday, October 19, 2018
20181011-Site Check -GLOCK
Check sample lines for condensation /No water
Check 2 charcoal for condensation
Pumps working fine
Out Flow pressure= 9 Psi
Cylinder Pressure =1600
No leaks in trailer
Check 2 charcoal for condensation
Pumps working fine
Out Flow pressure= 9 Psi
Cylinder Pressure =1600
No leaks in trailer
20181005-SP-Site Check - GLOCK
Check sample lines for condensation
Check 2 charcoal for condensation
Pumps working fine
OutFlow pressure= 10Psi
Cylinder Pressure =1600
Mosquitoes in the outside A/C
No leaks in trailer
Check 2 charcoal for condensation
Pumps working fine
OutFlow pressure= 10Psi
Cylinder Pressure =1600
Mosquitoes in the outside A/C
No leaks in trailer
Monday, October 8, 2018
20181008 SP Site Visit
Changed Teflon filter
Cleaned out dead mosquitoes from AC
NOx alarm: low pressure (189/200)
NOx cylinder out flow pressure 9psi
NOx cylinder pressure 1550psi
Cleaned out dead mosquitoes from AC
NOx alarm: low pressure (189/200)
NOx cylinder out flow pressure 9psi
NOx cylinder pressure 1550psi
Friday, September 28, 2018
Friday, September 21, 2018
Friday, August 24, 2018
Wednesday, August 8, 2018
20180802-Site Check and Calibratiom
Cylinder Pressure= 1600 psi
Out Flow Pressure= 12 psi
Droppings inside trailer but no animals visible
Need to replace charcoal scrubber
Condensation in Nox line
Replaced Teflon filter
Emptied #2 water trap
Out Flow Pressure= 12 psi
Droppings inside trailer but no animals visible
Need to replace charcoal scrubber
Condensation in Nox line
Replaced Teflon filter
Emptied #2 water trap
Thursday, June 28, 2018
Monday, June 18, 2018
Thursday, May 31, 2018
05222018 O3 troubleshooting
Installed new UPS
Replaced O3 instrument- we will troubleshoot cyclical noise in the lab
Small amount of water in trap 2
Replaced O3 instrument- we will troubleshoot cyclical noise in the lab
Small amount of water in trap 2
Wednesday, May 16, 2018
20180515 Site visit - Velasco
12psi 1650psi
NOx alarm pressure
No condensation in lines
Empty trap #2
Tried to figured out what is causing noise on the O3 instrument. Noise went away while I was at the site but it came back a few hours later.
Insulated lines to prevent condensation.
NOx alarm pressure
No condensation in lines
Empty trap #2
Tried to figured out what is causing noise on the O3 instrument. Noise went away while I was at the site but it came back a few hours later.
Insulated lines to prevent condensation.
Wednesday, May 9, 2018
Friday, April 20, 2018
20180418 - Site check - Velasco
| Name | Velasco |
| Date | 4/18/2018 |
| Trailer leaking | No |
| Trailer condition | Good |
| Pests | no |
| Ants | no |
| Alarms - O3 | No |
| Alarms NOx * | Flow/Pressure |
| Cylinder pressure (psi) | 1660 |
| Out flow pressure (psi) | 12 |
| Water in Water Trap** | no |
| Trailer Temperature | Normal |
| A/C Fan blowing | yes |
| Other | |
| Comments | Emptied water trap #2. Replaced Teflon filter. Sprayed ant insecticide. |
20180414 - Site check - Glock
| Name | Glock |
| Date | 4/14/2018 |
| Trailer leaking | no |
| Trailer condition | good |
| Pests | no |
| Ants | no |
| Alarms - O3 | no |
| Alarms NOx * | no |
| Cylinder pressure (psi) | 1700 |
| Out flow pressure (psi) | 12 |
| Water in Water Trap** | Yes |
| Trailer Temperature | Normal |
| A/C Fan blowing | yes |
| Other | |
| Comments | some condensation trap 2 |
20180409 - Site check - Glock
| Name | Glock |
| Date | 4/9/2018 |
| Trailer leaking | No |
| Trailer condition | Good |
| Pests | No |
| Ants | No |
| Alarms - O3 | |
| Alarms NOx * | |
| Cylinder pressure (psi) | 1700 |
| Out flow pressure (psi) | 12 |
| Water in Water Trap** | Yes |
| Trailer Temperature | Normal |
| A/C Fan blowing | Yes |
| Other | |
| Comments | some condensation trap # 2 |
20180331 - Site check - Glock
| Name | Glock |
| Date | 3/31/2018 |
| Trailer leaking | No |
| Trailer condition | Good |
| Pests | No |
| Ants | No |
| Alarms - O3 | |
| Alarms NOx * | |
| Cylinder pressure (psi) | 1700 |
| Out flow pressure (psi) | 12 |
| Water in Water Trap** | No |
| Trailer Temperature | Normal |
| A/C Fan blowing | Yes |
| Other | |
| Comments |
180326 Site visit
Name Velasco
Date 3/26/2018
Trailer leaking No
Trailer condition Good
Pests No
Ants No
Alarms - O3 Bench Temp high
Alarms NOx * Flow/Pressure
Cylinder pressure (psi) 1700
Out flow pressure (psi) 12
Water in Water Trap** No
Trailer Temperature Warm
A/C Fan blowing Yes
Other Comments: Site visit. Changed setting for AC low cooling fan at low cooling temp. too windy for ant killer spray.
Monday, March 26, 2018
20180320 Site deployment
NO cylinder 1700 psi - Out flow pressure = 12psi
Site deployed
Everything running.
Will spray for ants next time
Site deployed
Everything running.
Will spray for ants next time