Friday, June 28, 2019

Monday, June 24, 2019

20190621 Quarterly Cals

O3 and NOx quarterly calibrations
Checked sample lines for condensation or dirt. Removed condensation
Checked the canisters for condensation
Checked water traps. Emptied one that had water
Pumps running
No alarms present
Out flow- 12 psi
cylinder pressure - 1500
No pests or insects
A/C is working. Turned up the temperature
No water leaks
Replaced teflon filters
trouble shot iBoot

20190616 Site Visit

Tammy Edwards
Checked sample lines for condensation
Checked three water traps
Pumps are running
No alarms
Outflow pressure at 10 psi
Cylinder pressure at 1500
No pests or insects
A/C working
No water leaks

Monday, June 3, 2019

20190528 SP Troubleshoot - computer offline

Smith point computer went offline again. 
iBoot was troubleshooted since computer keeps turning off.  Had to re-work Ethernet cables
Keyboard batteries were leaking and had to be replaced with new ones
The weeds around trailer were cut by someone at the RV park.

20190520 SP site visit

Replaced Teflon filter
Small amount of condensation in water trap #2.  Condensation was removed
Condensation in the pump exhaust line near Carulite scrubber
Gas flow pressure ~11 psi.  Cylinder pressure 1500
No ant activity observed.  Too windy to reapply ant killer
Dropped AC Temperature a tiny bit to prevent condensation
AC was draining toward the door, tilted trailer to the front so it will drain that way.